4-11
IM CW500-01EN
Notes on Settings
If the current range is set to AUTO, only “Power + Harmonics” or “Power
only” can be selected with “1. Select desirable recording item.” If you want
to record power quality, change the current range to a fixed range before
starting to record. Only wiring and recording settings can be specified in
“Quick start guide.” Nominal voltage, nominal frequency, threshold for
power quality events (included in the measurement settings), and filter
coefficient (ramp) for flicker measurement must be set in advance. Set
these items from the SETUP menu. Note that the “+ Clamp” option clamp-
on probe setting is automatically set to OFF.
4.2 Starting to Record with “Quick start guide”