IM CA500-02EN
55
Thermocouple
Accuracy (1 year, t: measurement temperature)
Specification
Temperature Range
Measurement
Accuracy
±
[°C]
S
-20.0°C ≤ t < 0.0°C
1.4+|t-100| x 0.50%
IEC60584-1
1, 2
0.0°C ≤ t < +100.0°C
1.4+|t-100| x 0.50%
+100.0°C ≤ t ≤ +1768.0°C
1.4
B
+600.0°C ≤ t < +800.0°C
1.5
IEC60584-1
1, 2
+800.0°C ≤ t < +1000.0°C
1.2
+1000.0°C ≤ t ≤ +1820.0°C
1.1
C
0.0°C ≤ t < +1000.0°C
1.0
IEC60584-1
1
+1000.0°C ≤ t ≤ +2315.0°C
1.0+(t-1000) x 0.08%
XK
-200.0°C ≤ t < 0.0°C
0.6+|t| x 0.30%
GOST R
8.585-2001
0.0°C ≤ t < +300.0°C
0.6
+300.0°C ≤ t ≤ +800.0°C
0.7
A
0.0°C ≤ t < +1000.0°C
1.2
IEC60584-1
+1000.0°C ≤ t ≤ +2500.0°C
1.2+(t-1000) x 0.06%
D
(W3Re/W25Re)
0.0°C ≤ t < +300.0°C
1.8
ASTM E1751/
E1751M
+300.0°C ≤ t < +1500.0°C
1.2
+1500.0°C ≤ t ≤ +2315.0°C
2.2
G
(W/W26Re)
+100.0°C ≤ t < +300.0°C
1.8
ASTM E1751/
E1751M
+300.0°C ≤ t < +1500.0°C
1.2
+1500.0°C ≤ t ≤ +2315.0°C
2.2
PLATINEL II
0.0°C ≤ t < +100.0°C
1.8
ASTM E1751/
E1751M
+100.0°C ≤ t < +1000.0°C
1.8
+1000.0°C ≤ t ≤ +1395.0°C
2.2
PR20-40
0.0°C ≤ t < +500.0°C
11.0
ASTM E1751
+500.0°C ≤ t < +1000.0°C
4.0
+1000.0°C ≤ t ≤ +1888.0°C
2.0
Accuracy guaranteed under +23°C±5°C, 20 to 80%RH
Add a temperature coefficient of 0.05°C/°C for -10°C to +18°C and +28°C to
+50°C.
Display resolution of source/measured values: 0.1°C
Accuracy when using the included binding post (99045).
1 Complies also with JIS C 1602
2 The setting can be changed to comply with IPTS-68 (JIS C 1602 1981).
5. Specifications
Summary of Contents for CA550
Page 3: ......