5. Specifications
5.1 DC Voltage, DC Current, Resistance and Pulse
Source
DC Voltage Source
Range Resolution Source Range
Accuracy (1 year)
±(% of s offset)
Notes
CA500
CA550
100 mV
1 μV
±110.000 mV
0.015% + 10 μV 0.015% + 5 μV
Maximum output
current: 10 mA
1-5V
0.1 mV
0.0000 to
6.0000 V
0.015% + 0.5 mV
Maximum output
current: 10 mA
Value output function
supporting square
root computation is
available.
5 V
0.1 mV
±6.0000 V
0.015% + 0.5 mV
Maximum output
current: 10 mA
30 V
1 mV
±33.000 V
0.015% + 5 mV
Maximum output
current: 1 mA
Accuracy guaranteed under +23°C±5°C, 20 to 80%RH
Add a temperature coefficient of 0.005% of Range/°C for -10°C to +18°C and +28°C
to +50°C.
The source accuracy is defined for the following conditions.
• The specified cable is used to source the signal.
• The effect of the voltage drop in the cable is excluded.
• The effect of the aging of the instrument is excluded.
DC Current Source
Range Resolution Source Range
Accuracy (1 year)
±(% of s offset)
Notes
CA500
CA550
20 mA
1
μA
±24.000 mA
0.015% + 3 μA 0.010% + 2 μA
Source voltage: 0 to
+20 V
4-20 mA
1
μA
0.000 to
24.000 mA
(4.000 mA steps)
0.015% + 3 μA 0.010% + 2 μA
Source voltage: 0 to
+20 V
Value output function
supporting square
root computation is
available.
20 mA
SIMULATE
1
μA
0.000 to
+24.000 mA
0.015% + 3 μA 0.010% + 2 μA
External power
5 to +28 V
Accuracy guaranteed under +23°C±5°C, 20 to 80%RH
Add a temperature coefficient of 0.005% of Range/°C for -10°C to +18°C and +28°C
to +50°C.
Summary of Contents for CA550