Installation Conditions
Install the instrument in a place that meets the following conditions.
• Operating Altitude and Ambient Temperature and Humidity
• Use the instrument in the following environment.
• Ambient temperature
-10°C to 50°C
• Ambient humidity
80% RH or less for -10°C to 40°C, 50% RH or less
for over 40°C
No
condensation
• Operating altitude
Up to 2000 m
Note
• To ensure high accuracy, operate the instrument within 23°C ± 5°C.
• When using the instrument in a place where the ambient humidity is 30% or less, take
measures to prevent static electricity such as using an anti-static mat.
• Condensation may occur if the instrument is moved to another place where the ambient
temperature or humidity is higher, or if the temperature changes rapidly.
In such cases, allow the instrument adjust to the new environment for at least an hour
before using the instrument.
Install the instrument in a proper location by referring to “Safety Precautions” on
page 8.
Do not install the instrument in the following places, as it can cause incorrect
measurements or damage the instrument.
• In direct sunlight or near heat sources
• In an environment with excessive amounts of soot, steam, dust, or corrosive gas
• Near strong magnetic field sources
• Near noise sources, such as high-voltage equipment or power lines
• Where the level of mechanical vibration is high
• On an unstable surface
• Outdoors or in locations subject to rain or water