Between signal terminals and excitation current
terminal : 2000V AC for 1 minute or 2400V AC for
1 second.
Remote Flowtube (optional code WT1)
Between excitation current terminal and ground
terminal : 1000V AC for 1 minute
Remote Flowtube (optional code WT2)
Between excitation current terminal and ground
terminal : 1500V AC for 1 minute
Between signal terminals and excitation current
terminal : 1500V AC for 1 minute
CAUTION
*1: When performing the Insulation Resistance Test or the
Withstand Voltage Test, please obey the following
caution.
• Following the relevant test, wait for more than 10
seconds after the power supply has been turned off
before removing the cover.
• Remove all wires from terminals before testing.
• When the power terminal has a lighting protector
(optional code A), remove the short bar at the ground
terminal.
• After testing, be sure to discharge by using a resistance
and return all wires and the short bar to its correct
position.
• Screws must be tightened to a torque of 1.18 N-m or
more.
• After closing the cover, the power supply can be
restored.
Safety Requirement Standards:
EN61010-1
EN61010-2-030
• Altitude at installation site: Max. 2000 m above sea level
• Installation category based on IEC1010:
Overvoltage category II (“II” applies to electrical equip-
ment which is supplied from the fixed installation like
distribution board.)
• Pollution degree based on IEC1010
Pollution degree 2 (“Pollution degree” describes the
degree to which a solid, liquid, or gas which deteriorates
dielectric strength or surface resistivity is adhering. “2”
applies to a normal indoor atmosphere.)
EMC Conformity Standards:
,
EN61326-1 Class A, Table 2 (For use in industrial locations)
EN61326-2-3
EN61000-3-2 ClassA
EN61000-3-3
CAUTION
This instrument is a class A product, and it is designed for
use in the industrial environment.
Please use this instrument in the industrial environment
only.
■
NORMAL OPERATING CONDITIONS
Ambient Temperature:
–40 to +60
°
C (–40 to +140
°
F)
*1: Minimum temperature should also be limited according
to minimum fluid temperature of flow tube’s
specification.
Refer to description of “Fluid Temperature and
Pressure”.
*2: Indicator’s operating range (integral flowmeter): –20 to
+60
°
C (–4 to +140
°
F)
*3: Maximum temperature should be +50
°
C (+122
°
F) in the
case of power supply code 2 (integral flowmeter).
Ambient Humidity:
0 to 100%
Lengthy continuous operation at 95% or more is not
recommended.
Power Supply (integral flowmeter):
Power supply code 1:
• AC specifications
Rated power supply: 100 to 240 V AC, 50/60 Hz
(Operating voltage range: 80 to 264 V AC)
• DC specifications
Rated power supply: 100 to 120 V DC
(Operating voltage range: 90 to 130 V DC)
Power supply code 2:
• AC specifications
Rated power supply: 24 V AC, 50/60 Hz
(Operating voltage range: 20.4 to 28.8 V AC)
• DC specifications
Rated power supply: 24 V DC
(Operating voltage range: 20.4 to 28.8 V DC)
Supply Voltage and
Cable Length for Power Supply Code 2
0
20.4
22
24
26
28.8
200 ( 660)
446 (1460)
600 (1970)
796 (2610)
F01.EPS
Cable cross section area: 1.25 mm
2
Cable cross section area: 2 mm
2
Allo
w
a
b
le cab
le length m(ft)
Usable range E (V)
Fluid Conductivity:
Size 2.5 to 10 mm (0.1 to 0.4 in.): 5
µ
S/cm or larger
Size 15 to 125 mm (0.5 to 5 in.): 1
µ
S/cm or larger
Size 150 to 400 mm (6 to 16 in.): 3
µ
S/cm or larger
Note: Fluids with large flow noise (pure water, fluid with low
conductivity and low viscosity such as alcohol) cause
the output fluctuation. Be careful that it affects the
flow rate measurement. It is recommended to use the
CA capacitance magnetic flowmeter.