IM 01E20D01-01E
8-11
8. EXPLOSION PROTECTED TYPE INSTRUMENT
(h) Fasten the cable by tightening the clamp nut.
(i) Tighten the lock nut on the union cover.
(j) Connect the cable wires to each terminal.
F0809.EPS
Adapter body
Rubber packing
Washer
Packing box
Gland
Clamp ring
Clamp nut
Union coupling
Union cover
Wrench
O-ring
Lock nut
Wrench
Lock nut
Cable
Apply a non-
hardening sealant to
the threads for
waterproofing.
Figure 8.4.3
Installing Flameproof Packing Adapter
(2-2) Cable Wiring Using Flameproof Metal
Conduit
• A seal fitting must be installed near the terminal box
connection port for a sealed construction.
• Apply a non-hardening sealant to the threads of the
terminal box connection port, flexible metal conduit
and seal fitting for waterproofing.
• Every wiring port must apply the same wiring as
Figure 8.4.4.
Apply a blanking plug to the unused wiring port of a
24V power supply version when a four-wire cable is
used for both power input and signal output.
If the blanking plug is necessary, always purchase it
(parts number: G9330DP) from Yokogawa.
F0810.EPS
Non-hazardous
area
Hazardous
area
Flameproof
heavy-gauge
steel conduit
Tee
Drain plug
Seal fitting
Gas sealing
device
Flameproof flexible
metal conduit
After wiring, impregnate the
fitting with a compound to seal tubing.
Apply a non-hardening
sealant to the threads
of these fittings for
waterproofing
Apply a blanking
plug to the unused
wiring port
Figure 8.4.4
Typical Wiring Using Flameproof Metal
Conduit