![YOKOGAWA AQ6370D User Manual Download Page 140](http://html2.mh-extra.com/html/yokogawa/aq6370d/aq6370d_user-manual_3665131140.webp)
6-20
IM AQ6370D-01EN
6.7 Curve Fitting
Procedure
Setting the Target Trace
1.
Press
TRACE
. The soft key menu for trace settings appears.
2.
Press the
ACTIVE TRACE
soft key. Traces A through G appear in soft keys.
3.
Press the Trace G soft key. The Trace G setting menu is displayed.
4.
Press the
VIEW G DISP/BLANK
soft key and select
DISP
. Each time you press
the soft key
DISP
switches to
BLANK
, or viceversa.
5.
Press the
CALCULATE G
soft key.
6.
Press the
CURVE FIT
soft key. The menu for selecting the trace to curve fit is
displayed.
7.
To curve fit Trace A press the
G=CRV FIT A
soft key, to curve fit Trace B press
G=CRV FIT B
, and to curve fit Trace G, press
G=CRV FIT C
.