12-2
IM 01E30D01-01EN
<12. OUTLINE>
HART
Communication Signal:
HART communication signal (superimposed on 4 to
20 mA DC signals)
Note: HART is a registered trademark of the FieldComm
Group.
Conditions of Communication Line:
Supply Voltage: 20.6 to 42 V DC
Load Resistance: 250 to 600 Ω (including cable
resistance)
Read Figure 12.2.
HART Protocol Revision:
HART protocol revision can be selected from 5 or 7
when ordering. (“-J” only)
The protocol revision can be changed by user config
-
uration.
The HART protocol revision at the time of shipment
is shown by the last number at the serial number
column of the name plate.
Note: Protocol revision supported by HART configuration
tool must be the same or higher than that of AXR.
Selection of HART 5/ HART 7
Output Signal Code
-E
-J
Ordering
Information
–
Specify “5”
Specify “7”
HART Protocol
Revision
HART 5
HART 7
Selection
guide
Require-
ment for
HART 7
function-
ality
NO
YES
Be sure to
confirm the
protocol
revision of
the HART
configura
-
tion tool
shown in
Note 2.
Other
condi-
tions
Not availa-
ble to switch
to HART 7
protocol af-
ter delivery.
Available
to switch
to HART 7
protocol af-
ter delivery
by user-
configura
-
tion.
Available
to switch
to HART 5
protocol af-
ter delivery
by user-
configura
-
tion.
Remarks
Note 1
Note 2
Note 2
Note 1: “-E” is HART5 exclusive model and will be terminat-
ed. “-J” is recommended for HART communication.
Note 2: HART protocol revision for the device and HART
configuration tool
HART 7 communication is supported by FieldMate
R2.04 or later.
Protocol revision supported by HART
configuration tool
5
7
AXR, HART 5
Available
Available
AXR, HART 7
Not available
Available
Indicator (*2):
Full dot-matrix LCD (128 x 64 pixels)
Operational switch: 4 magnet switches (including
push switches)
Lightning Protector:
When optional code A is selected, the lightning pro-
tector is built into the power supply and digital output
terminals.
Protection:
General-purpose Use/TIIS Flameproof type:
IP66/IP67, Type 4X
Explosion proof type except for TIIS:
In the case of explosion proof type except TIIS,
read description of “Enclosure” in “HAZARDOUS
AREA CLASSIFICATION”.
Converter Coating:
Case and Cover: Corrosion-resistant coating
Coating Color: Mint-green paint (Munsell 5.6 BG
3.3/2.9 or its equivalent)
Converter Material:
Case and Cover: Aluminum alloy
Mounting/Shapes:
• Electrical Connection: ANSI 1/2 NPT female
ISO M20 x 1.5 female
JIS G1/2 female
• Direction of Electrical Connection: The direction can
be changed even after delivery.
• Terminal Connection: M4 size screw terminal
Grounding:
Grounding resistance of 100 Ω or less is necessary.
When the optional code A is selected, grounding
resistance of 10 Ω or less shall be required.
* In case of TIIS explosion proof type, read descrip-
tion of HAZARDOUS AREA CLASSIFICATION.
* For an explosion proof type except for TIIS, follow
the domestic electrical requirements in each coun-
try.
Functions
How to Set Parameters (*2):
Both of four push buttons and magnet switches can
be used for setting. The magnet switches can be
used to set parameters without opening the case
cover.
The magnet switches need operational magnets
(optional code BM). These are also available as part
number F9840PA.
Users can choose a language on indicator from Eng-
lish, Japanese, German, French, Italian, and Span-
ish. Parameters can also be set with the configuration
tool (Such as HHT (handheld terminal) or FieldMate,
etc.). The language for the HHT is English only.
Instantaneous Flow Rate/Totalized Value Display
Functions (for models with an indicator) (*2):
The full dot-matrix LCD enables user selections of
displays from one line to three lines for:
• Instantaneous flow rate
• Instantaneous flow rate (%)
• Instantaneous flow rate (bar graph)
• Current output value (mA)
• Totalized forward-direction flow rate
• Totalized reverse-direction flow rate
• Totalized differential flow rate
• Tag No.
• Results of electrode adhesion diagnosis
• Communication type