<11. MAINTENANCE>
11-7
IM 01E30D01-01EN
This section describes the procedure for Adhesion Diagnostic using the display unit switches. (For more details
regarding setting methods using these switches, read Chapter 5.)
The parameters that are relevant to the adhesion diagnostic function are as follows.
[K10:Adhesion Check]
Execution of adhesion diagnostic function
[K11:Adhesion Status]
Display of the adhesion diagnosis level
[K12:Adh Measure Value]
Displays the resistance value for adhesion diagnose
[K13:Adhesion Level1]
Setting the resistance value for adhesion diagnostic level1
[K14:Adhesion Level2]
Setting the resistance value for adhesion diagnostic level2
[K15:Adhesion Level3]
Setting the resistance value for adhesion diagnostic level3
[K16:Adhesion Level4]
Setting the resistance value for adhesion diagnostic level4
[K17:Adhesion Check Cycle]
Setting of execution time for adhesion diagnosis
For more details regarding these parameters, read Chapter 6.
The parameter for executing the adhesion diagnostic function is
K10:Adhesion Check
.
The parameter
K10:Adhesion Check
will be used in the following description.
Setting Mode
Major item
parameter
Setting mode
J:Aux
K:Diagnosis
M:Adjustment
K10:Adhesion Check
No Execution
No Execution
Execution
K10:Adhesion Check
No Execution
Execution
No Execution
Major item
Sub-item
selection (D)
K10:Adhesion Check
Execution
Sub-item
Parameter
Search Mode
Parameter
Replacement
(execute) Mode
Start:
Major Item
Parameter
Search Mode
K:Diagnosis
23:Empty Level
10:Adhesion Check
11:Adh Status
F1109.EPS
SET
SET
INC
SET
SET
Once in Setting Mode, use the
INC
switch to
move the cursor to
K: Diagnosis
.
Touch the
SET
switch to access Sub-item
Parameter Search Mode.
Upon selection of
K: Diagnosis
, the cursor will be
positioned at
K10: Adhesion Check
. (Sub-item
selection (D)).
Touch the
SET
switch to access Parameter
Replacement Mode.
Touch the
INC
switch to move the cursor to
“Execution”.
Touch the
SET
switch to select the “Execution”.
In order to request confirmation, the entire display
flashes on and off. Touch the
SET
switch once
again at this time to fix selection of the Adhesion
Diagnosis function.
NOTE
When no operations are carried out for 20
seconds in the flashing state, the system will
automatically return to the Sub-item Parameter
Search Mode.
When the operations except
SET
are carried
out, the parameter cannot be set.