<7. OPERATION VIA BRAIN TERMINAL (BT200)>
7-4
IM 01E30D01-01EN
7.2.2 The data update and upload/
download function of BT200
(1) The data update of BT200
When the following parameters are displayed, the
measured data is updated automatically every
seven seconds.
Item
Name (BRAIN)
Item
Name (BRAIN)
A10
FLOW RATE (%)
G43
ALM RECORD2
A20
FLOW RATE
G44
ALM TIME 2
A21
FLW RATE (mA)
G45
ALM RECORD3
A30
TOTAL
G46
ALM TIME 3
A31
REV TOTAL
G47
ALM RECORD4
A32
DIF TOTAL
G48
ALM TIME 4
C44
VELOCITY CHK
K11
ADH STATUS
D14
TL RATE CHK
K12
ADH MEAS VAL
E14
PLS RATE CHK
K20
EMPTY STATUS
G20
4-20 SYS ALM
K21
DC VOLTAGE A
G40
OPERATE TIME
K22
DC VOLTAGE B
G41
ALM RECORD1
P20
W PROTECT
G42
ALM TIME 1
P23
SOFT SEAL
(2) Upload/download function of BT200
Upload function is used when the parameters of
one AXR are copied to the BT200. And download
function is used when the parameters copied to the
BT200 are set to another AXR.
The targeted parameters for upload and download
are following.
Item
Name (BRAIN)
Item
Name (BRAIN)
C11/B20 FLOW DAMPING D10/B30 TOTAL UNIT
C30
SIZE UNIT
D11/B31 TOTAL SCALE
C31
NOMINAL SIZE
E10/B32 PULSE UNIT
C40/B21 FLOW UNIT
E11/B33 PULSE SCALE
C41/B22 TIME UNIT
H10/B40 DISP SELECT1
C42/B23 FLOW SPAN
H11/B41 DISP SELECT2
C43/B24 FLOW DECIMAL
H12/B42 DISP SELECT3
7.2.3 BT200 Screens & Flow Rate Data
Display
Use the following procedure to display flow rate
data on the BT200.
• The display of flow rate data is updated every
5 seconds.
When the BT200 is turned
on, the message “Please
wait ...” is displayed for
several seconds, and then
the screen on the left is
shown.
Press the key.
or
Press the key
to display initial data as
shown on the left.
Press the ( )or
key.
The tag number
specified upon
ordering is found
here.
The Flow Rate Screen
will be displayed.
A maximum of three items
of data can be displayed.
The or
key can be used to move
the cursor up and down.
(Note that the key
can be used to move the
cursor to the top of the
next screen.
Press the
( )
key
to return to Menu Screen
(Major Item Parameter
Search Mode).
F0708.ai
( )
The Menu Screen (Major
Item Parameter Search
Mode) will be displayed.
If the key is
pressed while A: DISPLAY
is selected by the highlight
bar, the Parameter Screen
(Sub-Item Parameter
Search Mode) will be
displayed.
ÐÐÐWELCOMEÐÐÐ
BRAIN Terminal
ID:BT200
Check cable connection
and press the Ent key.
UNTIL FEED
PARAMETER
01:MODEL
AXR
02:TAG NO
FI-1101
03:SELF CHECK
GOOD
OK
OK
MENU
A:DISPLAY
B:EASY SETUP
C:BASIC SETUP
D:TOTAL SET
E:PULSE SET
F:STATUS FUNC
HOME SET ADJ ESC
PARAMETER
A10:FLOW RATE(
%
)
80.0
%
A20:FLOW RATE
100m3/h
A21:FLW RATE(mA)
16.800mA
DATA DIAG PRNT ESC
ESC
ESC