Installation and service of the equipment should be performed by authorized and trained
personnel. Safety measures and local guidelines must be followed. Use good practices
wearing personal protective equipment (PPE) during the installation and service of the
equipment.
Electricity – Auto Pure lab water equipment should be powered by a grounded GFI electrical
connection. Do not open the system unless the system is unplugged. Before any installation
or service is performed on the equipment, make sure the power is isolated. To unplug the
system, do not pull directly on the cord but grip the plug and remove it from
the outlet.
Ultraviolet Light
– Auto Pure lab water systems may include an ultraviolet light lamp in a
stainless-steel housing. Make sure all power is removed prior to accessing the side panel to
perform any UV maintenance. Make sure gloves are used when handling the UV bulb.
(Gloves must be worn when handling the UV bulb. Which pr
otects the integrity of the bulb)
Do not look directly at the UV lamp while the system is operating.
Water Supply – Good piping practices should be followed feeding water to the lab equipment.
Isolation valves, pressure gauges and bleed valves are recommended on the main water
supply. Pressure should be regulated to meet the minimum and maximum specifications
stated for the equipment.
Sanitization – Make sure PPE is worn during any sanitization process. The Phoenix system
uses chlorine dioxide gas as a disinfectant. Gas is generated within the system, but the
chlorine dioxide gas will be discharged during sampling and flushing sequences. Chlorine
dioxide gas is a respiratory irritant. Avoid breathing fumes and ensure the area is sufficiently
ventilated.