All Models
Cartridge life is determined by a combination of elapsed time since installation and/or water
resistivity. The point at which the conditioning pack and polishing filter transition from green health
status to yellow can be adjusted.
See the table below for default health meter resistivity settings:
CARTRIDGE
RESISTIVITY
HEALTH METER COLOR
Conditioning
> 12 Meg
Ω
Green
Between 2 Meg
Ω
and 12 Meg
Ω
Yellow *
< 2 Meg
Ω
(default alarm set point)
Red
Polishing
> 16 Meg
Ω
Green
Between 10 MegΩ and 16 MegΩ
Yellow *
< 10
MegΩ (default alarm set point)
Red
*If set point is adjusted, (see Resistivity Alarm Setup), yellow health meter color will appear at the new set point.
The default elapse time for both the conditioning pack and polishing cartridge is six months. This
may be changed to 12-months in the advanced service menu. The health status meter will
automatically change from Green, to Yellow, to Red as time elapses even if the resistivity values
are above their alarm limit.
WC and WG SERIES:
UV bulb life is estimated at 1 year of normal use from date of installation. The UV bulb health
meter on the HOME SCREEN displays the proportional remaining life in this 1 year period as
shown below:
LIFE REMAINING
HEALTH METER COLOR
UV Bulb
> 2 months
Green
Between 1 month and 2 months
Yellow *
< 1 month
Red
31
HEALTH STATUS