SYMPTOM
CAUSE
REMEDY
Air is getting
trapped in the
system.
The point of use 0.2 micron filter
is bound with air.
Bleed the filter by opening the bleed valve and
allowing water to flow until all air is expelled.
The 0.05 micron filter is bound
with air. (WB & WG Series)
Bleed the filter by opening the bleed valve and
allowing water to flow until all air is expelled.
Filter Pack
Error
Cartridge not aligned
Reinstall cartridge and ensure that proximity
switch is engaged
Switch not working
If the error is still on while the cartridges are
engaged, the switch may need to be adjusted.
Contact Factory for instructions.
43
TROUBLESHOOTING