![background image](http://html.mh-extra.com/html/yamato/ba300/ba300_instruction-manual_905286018.webp)
15
4. OPERATION PROCEDURES
Constant Tempe Quick Auto Stop Mode
Quick auto stop
operation procedure
This mode is used to specify when unit should terminate constant
temperature operation. This mode is set during operation only.
1. Set timer during constant temperature operation
Confirm that unit is operating by confirming that FIXED TEMP
lamp is illuminated.
Press the TIMER key.
Top display will show
, indicating that the timer can be set.
Current timer setting will flash in center display.
Set the timer using the
▼▲ keys.
Timer function:
Maximum value for timer is "999 hours and 50 minutes ".
The time can be set in increments of one minute, under 99 hours
and 59 minutes.
Setting increments are 10 minutes, after 100 hours.
When
▼▲ keys are held down, values advance perpetually.
Press repeatedly for incremental adjustment.
2. Start quick auto stop operation
Press the START/STOP key for about one second after setting
timer.
FIXED TEMP and AUTO STOP lamps will illuminate, indicating
quick auto stop operation mode has started.
To manually stop/terminate quick auto stop operation
Operation stops automatically when timer reaches 0.00, and an
accompanying alarm sounds for approximately five seconds after
operation terminates.
Center display will show
, indicating end of operation, with
FIXED TEMP and AUTO STOP lamps illuminated. Press
START/STOP key to at any time during operation or after
operation ends, to terminate quick auto stop operation mode.
Displays will return to initial settings screen.
Editing or
confirming settings
Changing Temperature during operation is possible by pressing the
FIXED TEMP key. Press the ENTER key after changing temperature
setting.
Changing the timer setting during operation can be done without
terminating operation, simply by pressing the TIMER key. Note that
this will require adding additional time to the elapsed time. Press the
START/STOP key when changes have been entered, as desired.
Press the
▼ key at any time thereafter to see temperature setting,
operation mode and remaining time in the center display.
Summary of Contents for BA300