TX6n/5n/4n Reference Manual
18
Panel Operations
●
Parameters that can be edited
The parameters that can be edited will depend on the meter type.
* The setting range will differ depending on the meter type.
Parameter
Range
Explanation
Detect Enable
ON/OFF
Specifies whether alert detection will be enabled or disabled. An alert
message will be displayed if a point exceeding the specified value (Alert
Min/Max TH) is detected for the specified number of times (Detect
Count) within the specified duration (Detect Time). Detection is in 2 ms
intervals.
Alert Min TH
*
Specifies the lower threshold value for the alert.
Alert Max TH
*
Specifies the upper threshold value for the alert.
Detect Time
1–10s
Specifies the time during which points exceeding the threshold value will
be counted.
Detect Count
1–100
Specifies the number of times that a point exceeding the threshold value
must be counted in order to trigger an alert.
Peak Hold
ON/OFF
Turns peak hold on/off.
Unit
*
Specifies the display unit for the meter.
now
Time
Value
Max TH
Min TH
= Detect Count
Detect Time
Alert detection parameters