IOEM/REM Safey and Installation Guide
58
IOEM/REM Safey and Installation Guide
59
slave address 25/termination on
slave address 26/termination off
slave address 27/termination on
slave 28/termination on
slave 29/termination off
slave 30/termination on
slave 31/termination on
9.
10. Test network connections. Apply power to only the master and last slave
device on the network. If the master is not the first physical device on the
network, apply power to the master and the first and last physical slave
devices on the network See configure network menu in 5X00 user manuals
for testing procedure. Slave devices need to be tested one by one as they
are added to the network.
Step 7 Connect Communication Cable for Firmware
Update
An RS232 communication cable must be connected locally at the expansion module
to a PC running AquaManager in order to upload new IOEM and REM firmware.
Firmware is uploaded to IOEM and REM instruments using AquaManager. See
Updating Instrument Firmware at AquaManager help or follow the steps below to
upload new IOEM/REM firmware.
Note:
-
When making RS232 cable runs greater than 100 feet (30.5 m), use an
RS232 Line Extender/Booster. Mount the booster in an area where it is
protected from moisture.
Installation and Wiring
RS232 Communication
Use Category 5e patch cable from RS232 port at location
E
IOEM and
F
REM
RS232 Communication. Use an RJ45 to DB9 adapter to connect cable to PC 9 pin
com port.
IOEM com port
REM com port
Installation and Wiring