IOEM/REM Safey and Installation Guide
50
IOEM/REM Safey and Installation Guide
51
Installation and Wiring
WARNING: Disconnect external power to the unit before wiring.
AVERTISSEMENT : déconnectez l’alimentation externe de l’unité avant d’effectuer
un câblage quelconque.
1.
Perform steps 1 - 4 of IOEM and REM installation.
2.
3.
Strip five wires to terminate at network termination strip. Use appropriate
wire stripper to avoid wire nicks.
4.
Feed networking cable through drilled hole in rubber grommet and to
location
C
IOEM and
D
REM
.
Terminal strip is removable.
Note:
-
It is essential that wires be terminated cleanly. Nicks (stray wires, not
cleanly terminated) can result in unreliable network communication.
CAUTION
: Run high and low voltage wires/cables through separate bulk-
head and conduit.
IOEM network
REM network
5.
Terminate wires as described using category 5 cable: (color scheme may
be different than described below.)
-
Pair #1 wires: solid blue and white with blue stripe
-
Pair #2 wires: solid orange and white with orange stripe
-
1 wire green
-
Secure unused wires to avoid potential for shorting
Wire the master as follows
:
Wire Color
Terminal
solid orange
T (-)
white with orange stripe
T (+)
solid blue
R (-)
white with blue stripe
R (+)
green
Communications
Common
(C)
Wire all slaves as follows:
Wire Color
Terminal
solid blue
T (-)
white with blue stripe
T (+)
solid orange
R (-)
white with orange stripe
R (+)
green
Communications
Common
(C)
Installation and Wiring