Commissioning
Multi 3630 IDS
14
ba77170e03
10/2017
Operating condition
of the batteries
3.3.3 Switching on the meter
CAUTION
The batteries in the battery compartment are automatically
charged when the power pack is connected.
Make sure that only NiMH rechargeable batteries are in the
battery compartment. The charging process is optimized for
NiMH batteries. Other battery types can cause damage during
the charging process.
Make sure that the ambient temperature is not more than 40 °C
(104 °F) when the power pack is connected.
1
Connect the plug of the power pack to the socket for the power pack
on the Multi 3630 IDS.
2
Connect the original power pack to an easily accessible power outlet.
3
Charge the batteries completely prior to putting the meter into opera-
tion for the first time.
The charging process takes approx. 24 hours.
Symbol
Explanation
Line power operation with charge function
Batteries are automatically charged in the background
Battery operation
Batteries fully charged
Battery operation
Batteries almost empty
1
Press the
<On/Off>
key.
The meter performs a self-test.
The display shows the manufacturer's logo while the self-test is being
performed.
2
Connect the sensor.
The meter switches to the measuring mode (measured value display).
The meter has an energy saving feature to avoid unnecessary
power consumption during battery operation.
The energy saving feature switches off the meter during battery
operation if no key is pressed during the adjusted interval. (How to
set the switch-off interval, see section 11.6).
The switch-off interval of the energy saving feature is not active
when the meter is supplied with power via the power pack or the
USB-B cable.