Multi 3630 IDS
Conductivity
ba77170e03
10/2017
61
Calibration
evaluation
After calibration, the meter automatically evaluates the current status of the cal-
ibration. The evaluation appears on the display and in the calibration record.
Menu item
Setting/func-
tion
Explanation
Calibration
/
Calibration data
storage
/
Display
-
Displays the calibration records.
Further options:
Scroll through the calibration
records with
<
><
>
.
Output the displayed calibration
record to the interface with
<PRT>
.
Output all calibration records to
the interface with
<PRT_>
.
Quit the display with
<ESC>
or
<ENTER>
.
Switch directly to the measured
value display with
<M>
.
Calibration
/
Calibration data
storage
/
Output to
USB flash drive or
printer
-
Outputs the stored calibration data
to the USB-A interface (USB mem-
ory device/USB printer)
Calibration
/
Calibration data
storage
/
Output to
RS232/USB
-
Outputs the stored calibration data
to the USB-B interface (PC)
Display
Calibration
record
Cell constant [cm
-1
]
+++
Within the range
0.450 ... 0.500 cm
-1
Error
Error
Outside the range
0.450 ... 0.500 cm
-1
Eliminate the error according to chapter 15 W