1.2.3 IDS adapter for analog sensors
With the aid of an IDS adapter, you can also operate analog sensors
on the Multi 3630 IDS. The combination of the IDS adapter and analog sensor
behaves like an IDS sensor.
1.2.4 Automatic sensor recognition
The automatic sensor recognition for IDS sensors allows
to operate an IDS sensor with different meters without recalibrating
to operate different IDS sensors at one meter without recalibration
to assign measurement data to an IDS sensor
– Measurement datasets are always stored and output with the sensor
name and sensor series number.
to assign calibration data to an IDS sensor
– Calibration data and calibration history are always stored and output with
the sensor name and sensor series number.
to activate the correct cell constant for conductivity sensors automatically
to hide menus automatically that do not concern this sensor
To be able to use the automatic sensor recognition, a meter that supports the
automatic sensor recognition (e.g. Multi 3630 IDS) and a digital IDS sensor are
required.
In digital IDS sensors, sensor data are stored that clearly identify the sensor.
The sensor data are automatically taken over by the meter.
Information on available IDS adapters is given on the Internet.
Detailed information on the IDS adapter is given in the operating
manual of the adapter.