pH value
Multi 3630 IDS
34
ba77170e03
10/2017
5.2.4 Manual calibration (ConCal)
Make sure that in the sensor menu,
Buffer
menu, the
ConCal
buffer set is cor-
rectly selected (see section 11.1.1).
Use one to five buffer solutions in any order. The pH values of the buffer solu-
tions should differ by at least one pH unit.
19
Wait for the measurement with stability control to be completed or ter-
minate the stability control and take over the calibration value with
<ENTER>
.
The calibration display for the next buffer appears (voltage display).
20
If necessary, use
<M>
to finish the calibration.
The calibration record is displayed.
or
Switch to calibration with the next buffer with
<ENTER>
.
Calibration is automatically completed after the last buffer of a buf-
fer set has been measured. Then the calibration record is dis-
played.
The calibration line is determined by linear regression.
If single-point calibration was set in the menu, the calibration pro-
cedure is automatically finished with the measurement of buffer
solution 1 and the calibration record is displayed.
1
In the measured value display, select the measured parameter pH or
mV with
<M>
.
2
Start the calibration with
<CAL>
.
The calibration display for the first buffer appears (voltage display).
30.09.2015 08:00