Multi 3630 IDS
pH value
ba77170e03
10/2017
43
Carrying out a QSC
initial calibration
The QSC initial calibration is completed. The sensor is calibrated. If you want
to calibrate with special buffers for your measurements, you can subsequently
carry out a normal calibration with your buffers. The reference values deter-
mined with the QSC calibration are also used for the evaluation of normal cali-
brations. In the measured value display, the color scale of the QSC function is
always displayed. A double arrow on the color scale indicates the current sen-
sor evaluation.
As soon as the QSC function was enabled for an IDS sensor, it is
not possible to return to the sensor evaluation with the sensor sym-
bol for this sensor.
1
Open the menu for measurement settings with
<ENTER>
.
2
In the
QSC
menu, select
First calibration
with
<
><
>
.
The calibration display appears.
AutoCal QSC-Kit
is displayed as the
buffer.
Exclusively use the QSC-Kit for the QSC calibration. If you use other
buffers, you will have no valid QSC calibration.
3
Calibration with the buffers of the QSC-Kit is done like a normal three-
point calibration.
Follow the user guide.
Carry out the QSC initial calibration very carefully. It determines the
reference value for the sensor. This reference value cannot be
overwritten or reset.
As soon as the QSC function was enabled, it is not possible to
return to the sensor evaluation with the sensor symbol.
4
As soon as the three-point calibration has been successfully carried
out you can decide whether to accept or discard the calibration as the
QSC initial calibration.