Multi 3630 IDS
pH value
ba77170e03
10/2017
37
Continuing with
three- to five-point
calibration
16
Set the nominal buffer value for the measured temperature with
<
><
>
.
17
Accept the calibration value with
<ENTER>
.
The calibration display for the next buffer appears (voltage display).
18
If necessary, finish the calibration procedure as a two-point calibration
with
<M>
.
The calibration record is displayed.
15.03.2016 08:00
ConCal
19
Thoroughly rinse the IDS-pH sensor with deionized water.
20
Immerse the IDS-pH sensor in the next buffer solution.
21
When measuring without temperature sensor:
Measure the temperature of the buffer manually and enter it with
<
><
>
.
22
Start the measurement with
<ENTER>
.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.
23
Wait for the measurement with stability control to be completed or ter-
minate the stability control and take over the calibration value with
<ENTER>
.
The pH value of the buffer solution is displayed.
15.03.2016 08:00
ConCal