Multi 3410
Operation
76
ba75863y18
01/2015
Note
You can prematurely terminate the
Stability control
function manually
with
<MENU/ENTER>
at any time. If the
Stability control
function is pre-
maturely terminated, the current measurement data are output to the
interface without the AutoRead info.
Criteria for a stable
measured value
The
Stability control
function checks whether the measured values are
stable within the monitored time interval.
The minimum duration until a measured value is assessed as stable is
the monitored time interval. The actual duration is mostly longer.
3
Using
<MENU/ENTER>
, start a further measurement with
Sta-
bility control
.
or
Release the frozen measured value again with
<AR>
.
The display switches to the measured value display.
The [AR][HOLD] status display disappears.
Measured
parameter
Time interval
Stability in the time interval
Conductivity
ϰ
10 seconds
Δ
: Better than
1.0% of measured value
Temperature
10 seconds
Δ
: Better than 0.5 °C