Operation
Multi 3410
79
ba75863y18
01/2015
Type
Measuring cell used
Cal
Measuring cells whose cell con-
stant is determined by calibration
in the KCL control standard solu-
tion.
Calibration range:
0.450 to 0.500 cm
-1
The currently valid cell constant
is displayed in the status line.
man
Freely adjustable cell constant in
the range 0.450 ... 0.500 cm
-1
.
Man. cell const.
0.450 ...
0.475
...
0.500 cm
-1
The cell constant and be dis-
played and set here
Temp. comp. (TC)
/
Method
nLF
Lin
off
Procedure for temperature com-
pensation (see section 4.9.3).
This setting is only available for
the measured parameters,
ϰ
and
ρ
.
Temp. comp. (TC)
/
Linear coeff.
0.000 ...
2.000
...
10.000 %/K
Coefficient of the linear tempera-
ture compensation.
This setting is only available
when the linear temperature
compensation is set.
Temp. comp. (TC)
/
Reference temp.
20 °C
25 °C
Reference temperature
This setting is only available for
the measured parameters,
ϰ
and
ρ
.
TDS factor
0.40 ...
1.00
Factor for TDS value
Reset
-
Resets all sensor settings to the
delivery condition (see section
4.14.1)
Menu item
Possible
setting
Description
Summary of Contents for wtw Multi 3410
Page 1: ...OPERATING MANUAL ba75863e18 01 2015 Multi 3410 DIGITAL METER FOR DIGITAL IDS SENSORS...
Page 6: ...Multi 3410 6 ba75863y18 01 2015...
Page 14: ...Multi 3410 Overview 14 ba75863y18 01 2015...
Page 18: ...Multi 3410 Safety 18 ba75863y18 01 2015...
Page 106: ...Multi 3410 Operation 106 ba75863y18 01 2015...
Page 116: ...Multi 3410 What to do if 116 ba75863y18 01 2015...
Page 125: ......