Multi 3410
Operation
64
ba75863y18
01/2015
Stability control
(AutoRead )
The stability control function (AutoRead) continually checks the stability
of the measurement signal. The stability has a considerable impact on
the reproducibility of measured values. The display of the measured
parameter flashes until a stable measured value is available.
You can start a measurement with
Stability control
manually at any
time, irrespective of the setting for automatic
Stability control
tion 4.4.3) in the
System
menu.
Note
You can prematurely terminate the
Stability control
function manually
with
<MENU/ENTER>
at any time. If the
Stability control
function is pre-
maturely terminated, the current measurement data are not output to
the interface.
1
Freeze the measured value with
<AR>
.
The [HOLD] status indicator is displayed.
2
Using
<MENU/ENTER>
, activate the
Stability control
function
manually.
The [AR] status indicator appears while the measured value is
assessed as not stable. A progress bar is displayed and the
display of the measured parameter flashes.
The [HOLD][AR] status indicator appears as soon as a stable
measured value is recognized. The progress bar disappears
and the display of the measured parameter stops flashing.
The current measurement data is output to the interface. Mea-
surement data meeting the stability control criterion is marked
by AR.
3
Using
<MENU/ENTER>
, start a further measurement with
Sta-
bility control
.
or
Release the frozen measured value again with
<AR>
.
The display switches to the measured value display.
The [AR][HOLD] status display disappears.
Summary of Contents for wtw Multi 3410
Page 1: ...OPERATING MANUAL ba75863e18 01 2015 Multi 3410 DIGITAL METER FOR DIGITAL IDS SENSORS...
Page 6: ...Multi 3410 6 ba75863y18 01 2015...
Page 14: ...Multi 3410 Overview 14 ba75863y18 01 2015...
Page 18: ...Multi 3410 Safety 18 ba75863y18 01 2015...
Page 106: ...Multi 3410 Operation 106 ba75863y18 01 2015...
Page 116: ...Multi 3410 What to do if 116 ba75863y18 01 2015...
Page 125: ......