Note
You can display the slope in the units, mV/pH or % (see section 4.6.3).
Stability control
The calibration procedure automatically activates the stability control
function. The current measurement with stability control can be termi-
nated at any time (accepting the current value).
Calibration record
The new calibration values are displayed when a calibration procedure
is completed.
Display calibration data
and output to interface
You can have the data of the last calibration displayed (see section
4.6.8). Subsequently, you can transmit the displayed calibration data to
the interface, e. g. to a PC, with the
<PRT>
key.
Note
The calibration record is automatically transmitted to the interface after
calibrating.
Sample record
Determined
values
Displayed calibration data
1-point
Asy
Zero point =
Asy
Slope = Nernst slope
(-59.2 mV/pH at 25 °C)
2-point
Asy
Slp.
Zero point =
Asy
Slope =
Slp.
3-point to
5-point
Asy
Slp.
Zero point =
Asy
Slope =
Slp.
The calibration line is calculated
by linear regression.
Multi 3410
Ser. no. 09250023
CALIBRATION
pH
Calibration date
15.03.2014 16:13:33
SenTix 940
Ser. no. B092500013
TEC
Buffer 1
4.01
Buffer 2
7.00
Buffer 3
10.01
Voltage
1
184.0 mV 24.0 °C
Voltage
2
3.0 mV
24.0 °C
Voltage 3
-177.0 mV 24.0 °C
Slope
-60.2 mV/pH
Asymmetry
4.0 mV
Sensor
+++
etc...
Summary of Contents for wtw Multi 3410
Page 1: ...OPERATING MANUAL ba75863e18 01 2015 Multi 3410 DIGITAL METER FOR DIGITAL IDS SENSORS...
Page 6: ...Multi 3410 6 ba75863y18 01 2015...
Page 14: ...Multi 3410 Overview 14 ba75863y18 01 2015...
Page 18: ...Multi 3410 Safety 18 ba75863y18 01 2015...
Page 106: ...Multi 3410 Operation 106 ba75863y18 01 2015...
Page 116: ...Multi 3410 What to do if 116 ba75863y18 01 2015...
Page 125: ......