Operation
Multi 3410
47
ba75863y18
01/2015
Note
For
single-point calibration
, the instrument uses the Nernst slope
3
Thoroughly rinse the IDS sensor with deionized water.
4
Immerse the IDS pH sensor in buffer solution 1.
5
For measurements without temperature sensor
(e.g. when using an IDS adapter
)
:
Measure the temperature of the buffer manually and enter it with
<
><
>
.
6
Start the measurement with
<MENU/ENTER>
.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.
7
Wait for the end of the measurement with stability control or
accept the calibration value with
<MENU/ENTER>
.
The calibration display for the next buffer appears (voltage dis-
play).
8
If necessary, finish the calibration procedure as a single-point cal-
ibration with
<M>
.
The calibration record is displayed.
15.03.2014 08:00
15.03.2014 08:00