Multi 3620 IDS
What to do if...
ba77169d06
10/2021
99
15
What to do if...
15.1
General information
Sensor symbol
flashes
Display
Meter does not react
to keystroke
You want to know
which software
version is in the
meter or IDS sensor
Data transmission to
USB memory device
does not work
Cause
Remedy
– Calibration interval expired
– Recalibrate the measuring system
Cause
Remedy
– Batteries almost empty
– Charge the batteries (see section
)
– Replace the batteries (see section
Cause
Remedy
– Operating condition undefined
or EMC load unallowed
– Processor reset:
Press the
<MENU/ENTER>
and
<On/Off>
key simultaneously
Cause
Remedy
– E.g., a question by the service
department
– Switch on the meter.
– Open the menu,
<MENU/
ENTER_>
/
Storage & config
/
Sys-
tem
/
Service
. The instrument data
are displayed.
or
– Connect the sensor.
Press softkey [
Info
]/[
More
] The sen-
sor data are displayed (see section
4.1.6)
Cause
Remedy
– Connected USB memory device
was not recognized
– The USB memory device has
been formatted to a file system
which is not supported, e. g.
NTFS
– Use other USB memory device
– Reformat the USB memory device to
FAT 16 or FAT 32 file system
(Caution: Reformatting erases all
data on the USB memory device.
Back up all data before reformatting.)