Stability control
(AutoRead )
The stability control function (
AutoRead
) continually checks the stability of the
measurement signal. The stability has a considerable impact on the reproduc-
ibility of measured values. The display of the measured parameter flashes until
a stable measured value is available.
You can start the
Stability control
function manually at any time, irrespective of
the setting for automatic
Stability control
(see section 11.6.3) in the
System
menu.
Criteria for a stable
measured value
The
Stability control
function checks whether the measured values are stable
If the HOLD function is active, you can, e.g. start a manual mea-
surement with stability control.
2.
Release the frozen measured value again with
<AR>
.
The HOLD function is switched off.
The [HOLD] status display disappears.
1.
Freeze the measured value with
<AR>
.
The [HOLD] status indicator is displayed.
2.
Using
<MENU/ENTER>
, activate the
Stability control
function manu-
ally.
The [AR] status indicator appears while the measured value is
assessed as not stable. A progress bar is displayed and the display of
the measured parameter flashes.
The [HOLD][AR] status indicator appears as soon as a stable mea-
sured value is recognized. The progress bar disappears and the dis-
play of the measured parameter stops flashing.
The current measurement data is output to the interface. Measurement
data meeting the stability control criterion is marked by AR.
You can prematurely terminate the
Stability control
function manu-
ally with
<MENU/ENTER>
at any time. If the
Stability control
func-
tion is prematurely terminated, the current measurement data are
output to the interface without the AutoRead info.
3.
Using
<MENU/ENTER>
, start a further measurement with
Stability
control
.
or
Release the frozen measured value again with
<AR>
.
The display switches to the measured value display.
The [AR][HOLD] status display disappears.