Calibration evalua-
tion
After calibrating, the meter automatically evaluates the calibration. The zero
point and slope are evaluated separately. The worse evaluation of both is taken
into account. The evaluation appears on the display and in the calibration
record.
Menu item
Setting/func-
tion
Explanation
Calibration
/
Calibration data stor-
age
/
Display
-
Displays the calibration records.
Further options:
Scroll through the calibration
records with
<
><
>
.
Output the displayed calibra-
tion record to the interface with
<PRT>
.
Output all calibration records
to the interface with
<PRT_>
.
Quit the display with
<ESC>
or
<MENU/ENTER>
.
Switch directly to the mea-
sured value display with
<M>
.
Calibration
/
Calibration data stor-
age
/
Output to USB
flash drive or printer
-
Outputs the stored calibration
data to the USB-A interface (
USB
Host
) (USB memory device/USB
printer)
Calibration
/
Calibration data stor-
age
/
Output to
RS232/USB
-
Outputs the stored calibration
data to the USB-B interface (
USB
Device
) (PC)
Display
Calibration
record
Zero point
[mV]
Slope [mV/
pH]
+++
-15 ... +15
-60,5 ... -58
++
-20 ... +20
-58 ... -57
+
-25 ... +25
-61 ... -60.5
or
-57 ... -56
-
-30 ... +30
-62 ... -61
or
-56 ... -50
Clean the IDS sensor according to the
sensor operating manual