Multi 3620 IDS
Dissolved oxygen
ba77169d06
10/2021
53
Stability control
(AutoRead)
In calibration, the Stability control function (AutoRead) is automatically acti-
vated.
7.3.5 Calibration data
The calibration data can be displayed and then output to the interface.
Displaying the cali-
bration record
The calibration protocol of the last calibration is available in the menu
Calibra-
tion
/
Calibration record
. To open it in the measured value display, press the
<CAL_>
key.
The calibration records of the last 10 calibrations are available in the menu
Cali-
bration
/
Calibration data storage
/
Display
. To open the
Calibration
menu in the
measured value display, press the
<MENU/ENTER>
key.
3.
Start the calibration with
<CAL>
.
The last calibration data (relative slope) is displayed.
4.
Start the measurement with
<MENU/ENTER>
.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.
5.
Wait for the AutoRead measurement to be completed (status indicator
[HOLD][AR]) or
accept the calibration value with
<MENU/ENTER>
.
The calibration record is displayed and output to the interface.
6.
Switch to the measured value display with
<MENU/ENTER>
.
15.03.2014 08:00
The calibration record is automatically transmitted to the interface
after calibrating.