Dissolved oxygen
Multi 3620 IDS
54
ba77169d06
10/2021
Calibration evalua-
tion
After calibrating, the meter automatically evaluates the current status of the cal-
ibration. The evaluation appears on the display and in the calibration record.
Calibration evalua-
tion FDO
®
925
Menu item
Setting/func-
tion
Explanation
Calibration
/
Calibration data stor-
age
/
Display
-
Displays the calibration records.
Further options:
Scroll through the calibration
records with
<
><
>
.
Output the displayed calibra-
tion record to the interface with
<PRT>
.
Output all calibration records
to the interface with
<PRT_>
.
Quit the display with
<ESC>
or
<MENU/ENTER>
.
Switch directly to the mea-
sured value display with
<M>
.
Calibration
/
Calibration data stor-
age
/
Output to USB
flash drive or printer
-
Outputs the stored calibration
data to the USB-A interface (
USB
Host
) (USB memory device/USB
printer)
Calibration
/
Calibration data stor-
age
/
Output to
RS232/USB
-
Outputs the stored calibration
data to the USB-B interface (
USB
Device
) (PC)
Display
Calibration
record
Relative slope
+++
S = 0.94 ... 1.06
++
S = 0.92 ... 0.94 or
S = 1.06 ... 1.08
+
S = 0.90 ... 0.92 or
S = 1.08 ... 1.10
Error
Error
S < 0.90 or
S > 1.10
Perform error elimination accord-
ing to chapter 15 W