August 2020 – TD 316 OPERATING MANUAL – Motus Wave Sensor 5729
Page 55
3.10.7
Advanced System Output
Figure 3-31: Advanced System Output in User Maintenance
The properties in this section may be set to one of the tree options:
Off, Storage
and
Storage
.
Off
:
Parameter is turned off
Storage
:
Parameter is stored in logger but not presented on real-time output
Storage
:
Parameter is both stored in logger and presented through the real-time output
The parameters in this section are useful parameters for troubleshooting purpose.
Processing Time Output
is the time
used for processing all the data from the last wave integration time.
FE State Parameters Output
is an advanced system output which can be enabled to give more information about the
internal coprocessor and communication between the host processor and the coprocessor. Two output parameters,
FE
State
and
FE Notifications
are controlled by this setting. The
FE State
parameter
is 0 if everything is ok. The
FE
Notifications
parameter gives information about the last communications with the coprocessor.
HW State Parameter Output
is an advanced system setting which can be enabled to give more information about the
internal electronic hardware. The output is 0 if everything is ok.