Section 5: System Errors
5-4 | 14/11/2018
PHET_OP_GB Version 7© Xstrahl Ltd. All Rights Reserved
Error Code Table
No.
Message Displayed
Description
42
Less than 5 kV reached
Generator error. Displayed when the
generator fails to reach 5 kV within a defined
period of time after receiving an X-Ray ON
command. Latching error.
43
Interlock to Generator Lost
Generator error. Displayed when an exposure
is terminated by the removal of the interlock
input to the generator. Latching error
45
Prohibited Exposure
Displayed when the time entered in a
Time
treatment exceeds the set time limit. This error
will persist until a compliant time is entered.
(can be caused in a
Dose
treatment)
46
Uninitialized Filter
Displayed when an uninitialized
filter
is
selected or zero values detected for any of the
following: kV, mA, HVL, Type HVL, Dimension
or encoding.
47
Bad Applicator Data
Displayed when an uninitialized applicator is
selected; non-valid shape or zero values in
any of the following: width, breadth for
rectangular applicator, length or encoding.
48
Bad Dose Calibration
Displayed when a
filter
is selected for a dose
exposure with a non-valid reference applicator
or zero values for counts/second or counts/
MU.
49
Exposure Not Entered
Displayed when the X-Ray button is pressed
and the internal NEEDS status indicates the
exposure
is not set. This error will persist until
another treatment is set.
50
No Filter Selected
Displayed when the X-Ray button is pressed
and the internal NEEDS status indicates the
filter
is not set. This error will persist until
another treatment is set.
51
No Applicator Chosen
Displayed when the X-Ray button is pressed
and the internal NEEDS status indicates the
applicator
is not set. This error will persist until
another treatment is set.
52
Bad Temperature
Displayed when the temperature (in degrees
centrigrade) read from the transducer is less
than a
minimum
(10) or greater than a
maximum
(35).
Dose systems only
. This error
will persist whilst the temperature is invalid.
53
Bad Pressure
Displayed when the pressure (in milli-bars)
read from the transducer is less than a
minimum
(700) or greater than a
maximum
(1200).
Dose systems only
. This error will
persist whilst the pressure is invalid.
55
X-Ray Off Signal Open to
Power Up
Displayed if the X-Ray OFF button is open-
circuit at power up.
56
Shutdown by Safety1
Displayed if the X-Ray beam is stopped by the
door contact being open. Latching error
requires a new X-Ray ON signal to be reset.
Summary of Contents for RADiant
Page 1: ...RADiant Therapy System Operator Manual Language English...
Page 2: ......
Page 14: ...RADiant Operator Manual ii 14 11 2018 PHET_OP_GB Version 7 Xstrahl Ltd All Rights Reserved...
Page 16: ...RADiant Operator Manual iv 14 11 2018 PHET_OP_GB Version 7 Xstrahl Ltd All Rights Reserved...
Page 36: ...RADiant Operator Manual 4 2 14 11 2018 PHET_OP_GB Version 7 Xstrahl Ltd All Rights Reserved...
Page 45: ......