PHET_OP_GB Version 7 ©2017 Xstrahl Ltd. All rights reserved
12/02/18 | 5-1
Section 5:
System Errors
5
System Errors
Xstrahl system errors are displayed on the bottom right side of the exposure
screens in Concerto and Fisica.
Types of Errors
Normal Errors
Will persist whilst the problem exists; for example, a filter encoding error
will remain until the correct filter is fitted.
Latching Errors
These will persist after the problem goes away, but require further action
to clear. Pressing the X-Ray On button will clear this type of error.
Fatal Errors
Errors which cannot be cleared without turning the TP2 power off and
then back on at the isolator.
Faults
Error classification similar to Fatal errors and identifies if they are a
latching error
Summary of Contents for RADiant
Page 1: ...RADiant Therapy System Operator Manual Language English...
Page 2: ......
Page 14: ...RADiant Operator Manual ii 14 11 2018 PHET_OP_GB Version 7 Xstrahl Ltd All Rights Reserved...
Page 16: ...RADiant Operator Manual iv 14 11 2018 PHET_OP_GB Version 7 Xstrahl Ltd All Rights Reserved...
Page 36: ...RADiant Operator Manual 4 2 14 11 2018 PHET_OP_GB Version 7 Xstrahl Ltd All Rights Reserved...
Page 45: ......