Section 5: System Errors
5-8 | 14/11/2018
PHET_OP_GB Version 7© Xstrahl Ltd. All Rights Reserved
Error Code Table
No.
Message Displayed
Description
121
High Dose Rate Error
Displayed when the dose rate monitored in a
Dose
treatment is greater than the calibrated
dose rate by more than 3%. The error will
terminate X-Rays. The error requires the
control pod key to be counter-rotated to
standby
(position 2) and then back to
HT
(position 3) to clear before the X-Rays are
enabled.
122
Applicator Encoding Error
Displayed when the
applicator
bits do not
match those of the specified applicator.
123
Filter Encoding Error
Displayed when the
filter
bits do not match
those of the specified filter.
124
Generator Not Stopping
Displayed if the generator does not turn OFF
within two seconds.
125
Generator Not Starting
Displayed if the generator does not turn ON
within two seconds.
127
Generator Not Setup
Displayed if the generator cannot be set to the
required kV or mA. This will prevent X-Rays
ON.
129
Maintenance Due
Generator error. Displayed if the current date
exceeds the date chosen for routine
maintenance. Will not inhibit X-Rays ON.
193
Generator Interlock Open
Displayed when the interlock to the generator
is open; suppressed unless the interlock relay
is enabled.
195
PC Data Interlock
Displayed when the TP2 fails to receive a
software interlock from the PC at less than one
second intervals with the control pod key in
HT
(position 3).
196
Door Open
Displayed when the first safety signal is not
sensed at the TP2. This error will disconnect
the
safety contactor
.
197
Room Interlock Open
Displayed when the second safety signal is
not sensed at the TP2. This error will
disconnect the
safety contactor
.
228
Watchdog Failure
Displayed when the TP2 software fails to re-
trigger the watchdog in time. The error is not
resettable.
Fatal error.
229
Background Lockup
Displayed when the TP2 background software
fails to execute in time. The error is not
resettable.
Fatal error.
234
Other Trap
Displayed if an unexpected fault occurs. The
error is not resettable.
Fatal error
236
Bad Code Executed
Displayed if a bad instruction is detected. The
error is not resettable.
Fatal error
238
No Real Time Clock
NOT USED
243
Processor Clock Fault
Displayed if an internal processor clock fault
arises. The error is not resettable.
Fatal error.
Summary of Contents for RADiant
Page 1: ...RADiant Therapy System Operator Manual Language English...
Page 2: ......
Page 14: ...RADiant Operator Manual ii 14 11 2018 PHET_OP_GB Version 7 Xstrahl Ltd All Rights Reserved...
Page 16: ...RADiant Operator Manual iv 14 11 2018 PHET_OP_GB Version 7 Xstrahl Ltd All Rights Reserved...
Page 36: ...RADiant Operator Manual 4 2 14 11 2018 PHET_OP_GB Version 7 Xstrahl Ltd All Rights Reserved...
Page 45: ......