Section 5: System Errors
PHET_OP_GB Version 7 ©2017 Xstrahl Ltd. All rights reserved
14/11/2018| 5-3
Error Code Table
No.
Message Displayed
Description
23
Contactor Dropout
Displayed because the user has pressed the X-Ray
ON button too quickly. Wait for the X-Ray ON
button to illuminate green, and for the PC screen to
show Ready for an exposure, before pressing the
X-Ray ON button.
24
Residual kV too High
Generator error. Displayed if the generator kV
exceeds a certain value in X-Ray OFF.
25
Residual mA too High
Generator error. Displayed if the generator mA
exceeds a certain value in X-Ray OFF.
27
Communication Delay
Displayed if the communications from the
generator fails. This will inhibit or terminate
X-Rays ON.
28
All Filters in Box
Displayed when all
filters
are in the wall box
when a
filter
is selected (only certain systems).
Non-latching error; will reset automatically.
29
Two Filters Removed
Displayed when more than one
filter
is missing
from the wall box when a
filter
is selected (only
certain systems). Non-latching error; will reset
automatically.
30
Please Fit Filter
Displayed when no
filter
is fitted to the
sub-tube assembly when a
filter
is selected.
Non-latching error; will reset automatically.
34
Bipolar kV Matching Failure
Generator error. Displayed in bipolar systems
when the anode and cathode kV
measurements differ by a defined amount.
Latching error.
35
Bipolar mA Matching
Failure
Generator error. Displayed in bipolar systems
when the anode and cathode mA
measurements differ by a defined amount.
Latching error.
36
Feedback Open Circuit
Generator error. Displayed whilst the
connection to the control PCB from the
filament feedback PCB is open circuit
(CP Models).
37
Heatseat Sink Temperature
Generator error. Displayed whilst the heatseat
sink thermostat indicates a high temperature.
38
High Calibration Error
Generator error. Displayed when the value of
calibration voltage entered in an initializing
process is too high.
39
Low Calibration Error
Generator error. Displayed when the value of
calibration voltage entered in an initializing
process is too low.
40
Anode Over mA
Generator error. Displayed in bipolar systems
when the anode tank mA exceeds the desired
value. Latching error.
41
Anode Under mA
Generator error. Displayed in bipolar systems
when the anode tank mA fails to reach the
desired value within a defined period of time.
Latching error.
Summary of Contents for RADiant
Page 1: ...RADiant Therapy System Operator Manual Language English...
Page 2: ......
Page 14: ...RADiant Operator Manual ii 14 11 2018 PHET_OP_GB Version 7 Xstrahl Ltd All Rights Reserved...
Page 16: ...RADiant Operator Manual iv 14 11 2018 PHET_OP_GB Version 7 Xstrahl Ltd All Rights Reserved...
Page 36: ...RADiant Operator Manual 4 2 14 11 2018 PHET_OP_GB Version 7 Xstrahl Ltd All Rights Reserved...
Page 45: ......