unauthorized modifications be made to the systems. Follow this User Guide when setting-up and operating the
system.
Emissions Test
Compliance Electromagnetic Environment – Guidance
RF Emissions
CISPR 11
Group 1
The ForeSite IS System uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
RF Emissions
CISPR 11
Class B
The ForeSite IS System is suitable for use in all establishments except
domestic. However, it may be used in domestic establishments and those
directly connected to the public low voltage power supply network that
supplies buildings used for domestic purposes if the following warning is
heeded:
Warning: This equipment/ system is intended for use by healthcare
professionals only. This equipment/system may cause radio interference
or may disrupt the operation of nearby equipment. It may be necessary to
take mitigation measures, such as re-orienting or relocating the ForeSite
IS System or shielding the location.
Harmonic
Emissions IEC
61000-3-2
Class B
Voltage
Fluctuations/
Flicker Emissions
IEC 61000-3-3
Complies
IEC 60601-1-2:2014
The ForeSite IS System is intended for use in the electromagnetic environment specified below. The
customer or the user of the ForeSite IS System should assure that it is used in such an environment.
Immunity Test
IEC 60601
Test
Level
Compliance
Level
Electromagnetic Environment – Guidance
Electrostatic
Discharge (ESD) IEC
61000-4-2
±8 kV
Contact
±15 kV Air
±8 kV Contact
±15 kV Air
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the
relative humidity should be at least 30%.
Electrical Fast
Transient/Burst IEC
61000-4-4
±2 kV for
power
supply lines
±2 kV for
power supply
lines
Mains power quality should be that of a typical
commercial or hospital environment.
Surge
IEC 61000-4-5
±1 kV line(s)
to line(s)
±2 kV line(s)
to earth
±1 kV line(s) to
line(s)
±2 kV line(s) to
earth
Mains power quality should be that of a typical
commercial or hospital environment.
Power Frequency
Magnetic Field
(50/60 Hz)
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at
levels characteristic of a typical location in a typical
commercial or hospital environment.