ForeSite Intelligent Surface System User Guide
Maintenance and Cleaning
XSENSOR Technology Corporation
28 of 46
Cleaning and Disinfecting the Sensor Surface
Thoroughly clean the ForeSite IS System according to the instructions in this section.
As outlined below, the sensor surface must be cleaned with two bleach-wipe-rinse cycles.
Remove the cotton sheet covering the system prior to performing the cleaning procedure.
Prepare a bleach solution by diluting common household bleach at a 1:10 dilution.
H
2
O
Using a soft bristled brush that has been dipped in the prepared bleach solution, remove all visual
evidence of gross soil.
Rinse off residual cleaning solution using a clean soft cloth that has been wetted with cool tap water by
wiping until all evidence of residual cleaning solution is removed.
2
O
H
1
2
3