Compliance with Emission and Immunity
Standards
The equipment generates (and can radiate) radio-frequency energy. It can cause interference with adjacent
equipment if it is not installed and used in accordance with the instructions in this manual. Even with adequate
precautions, there is no guarantee that interference will not occur. To determine if the ForeSite IS System
causes interference, move the system away from adjacent equipment.
Portable and mobile RF communications equipment can affect the operation of electrical equipment.
Precautions should be taken to minimize the potential of this occurrence. Such precautions include, but are not
limited to, separation of the RF communication equipment and the ForeSite IS System.
Standards IEC 60601-1-2:2014
ANSI C63.4:2014
American National Standard for Methods of Measurement of Radio-Noise Emissions
from Low- Voltage Electrical and Electronic Equipment in the Range of 9 kHz to 40
GHz
IEC
60601-1-2:2014
Medical Electrical Equipment - Electromagnetic Compatibility
CISPR
11:2009+A1:2010
Industrial, scientific and medical equipment - Radio- frequency disturbance
characteristics
IEC
61000-3-2:2014
Limits for Harmonic Current Emissions (equipment input current ≤ 16A per phase)
IEC
61000-3-3:2013
Limitation of Voltage Changes, Voltage Fluctuations and Flicker in Public Low-Voltage
Supply Systems, for equipment with rated current ≤ 16A per phase and not subject to
conditional connection
IEC
61000-4-2:2008
Testing and Measurement Techniques - Electrostatic Discharge Immunity Test
IEC
61000-4-3:2010
Testing and Measurement Techniques - Radiated, Radio-Frequency, Electromagnetic
Field Immunity Test
IEC
61000-4-4:2012
Testing and Measurement Techniques - Electrical Fast Transient/Burst Immunity
Test
IEC
61000-4-5:2014
Testing and Measurement Techniques - Surge Immunity Test
IEC
61000-4-6:2006
Testing and Measurement Techniques - Immunity to Conducted Disturbances,
Induced by Radio- Frequency Fields
IEC
61000-4-8:2009
Testing and Measurement Techniques - Power Frequency Magnetic Field Immunity
Test
IEC
61000-4-11:2004
Testing and Measurement Techniques - Voltage Dips, Short Interruptions and Voltage
Variations Immunity Tests
ISO 17025
General requirements for the competence of testing and calibration laboratories
*XSENSOR declares that the ForeSite IS complies with the European Communities Council Medical Device
Directive 93/42/EEC amended by 2007/47/EC. This declaration will be deemed invalid should any