Using the External Shutdown Function
M370046-01
3-31
3
Defining the Polarity of the External Shutdown Signal
1. Turn the 9-position mode control to the PRT position or press the
rotary Adjust/Enter control if the control knob is already at the PRT
position.
PrO OUP
is displayed on the output voltage display.
2. Turn the rotary Adjust/Enter control until
PrO SD
is displayed.
3. Press the rotary Adjust/Enter control to display
SD
0
(shutdown on
logic 0) or
SD1
(shutdown on logic 1).
4. Press the rotary Adjust/Enter control to commit the desired polarity
setting.
For example: If you set the shutdown logic to 1 (
SD1
) and apply a
logic high (greater than 1.2 Vdc) to pin J3.12 relative to pin J3.6 on
the ISOL connector, the power supply will shut down. Conversely, if
you set the shutdown logic to 0 (
SD0
), the power supply will only
operate when you have a logic high (greater than 1.2 Vdc) to pin J3.2
relative to pin J3.6.
The SCPI command (s) for these instructions are:
[:]OUTPut[<channel>]:POLarity
Important:
There is an internal pullup resistor inside the shutdown circuit so
the shutdown polarity should be set to 0 if it is not being used.
Summary of Contents for XTR 850 Watt
Page 2: ......
Page 3: ...XTR 850 Watt Series Programmable DC Power Supply firmware v1 09 and below Operating Manual...
Page 6: ......
Page 18: ...xvi...
Page 30: ...1 8...
Page 52: ...Local Operation 3 8 M370046 01 Figure 3 3 Front Panel Menu System...
Page 126: ...Analog Programming APG and Isolated Analog Programming ISOL 4 34 M370046 01...
Page 198: ...5 72...
Page 234: ...A 36...
Page 284: ...IX 4...
Page 285: ......