Calibration and Troubleshooting
B-16
M370046-01
Calibration Procedure for Isolated Modes
The main output and the non-isolated mode must be calibrated first.
Isolated Voltage Monitoring Calibration
To calibrate the isolated voltage monitoring:
1. Disconnect the load and connect the voltmeter to the output.
2. Connect the voltmeter to the isolated voltage monitoring output lines
(J3.5 - J3.6).
3. Set the Voltage APG level to 4 V. The SCPI command is:
SYST:COMM:APR:LEV 4
4. Set the input source to local for voltage analog programming. The
SCPI command is:
SYST:REM:SOUR LOC
5. Set the main output voltage to 100% by sending the command:
SOUR:VOLT <ModelVolt>
or use the rotary adjust/Enter control.
6. Turn the main output to On by pressing the OUTPUT ENABLE Main
button on the front panel or send the command:
:OUTP ON
The OUTPUT ENABLE Main button will illuminate.
7. Enter the interactive calibration mode by sending the SCPI command:
CAL:MON:ISOL
8. Press the terminal keys – or + to reduce or increase the monitoring
value. Adjust the monitoring value displayed on the voltmeter that
was setup in Step 1 to read as close as possible to 4.000 V.
9. Press any key other than the + or – keys to exit from interactive
calibration mode.
Important:
For maximum accuracy at a specific APG level. Follow the
calibration procedure again but use the intended operating voltage level in steps 3
and 8.
Summary of Contents for XTR 850 Watt
Page 2: ......
Page 3: ...XTR 850 Watt Series Programmable DC Power Supply firmware v1 09 and below Operating Manual...
Page 6: ......
Page 18: ...xvi...
Page 30: ...1 8...
Page 52: ...Local Operation 3 8 M370046 01 Figure 3 3 Front Panel Menu System...
Page 126: ...Analog Programming APG and Isolated Analog Programming ISOL 4 34 M370046 01...
Page 198: ...5 72...
Page 234: ...A 36...
Page 284: ...IX 4...
Page 285: ......