Analog Programming Mode
M370046-01
4-13
4
Query for Analog Current Input Level
To query for analog current input level from non-isolated input:
1. Turn the 9-position mode control to the CAP position or press the
rotary Adjust/Enter control if the mode control is already at the CAP
position.
CAPr
is displayed on the output voltage display.
2. If the mode is currently
nonE
then select any other APG mode. If this
step is done, be sure to restore the APG setting to
nonE
when the
operation is complete. Otherwise, do not rotate the rotary Adjust/
Enter control to keep the current APG mode.
3. Press the rotary Adjust/Enter control to select the APG mode.
The output voltage display should read
LE C
and the output current
display will display APG level.
4. Press the rotary Adjust/Enter control to return to normal mode.
5. If the APG mode was modified in step 2, execute the procedure to
deactivate the APG voltage mode.
Quick Tip
Remote operation
uses these SCPI
commands. See
page 5–1.
The SCPI command (s) for these instructions are:
[:]SYSTem[<channel>]:REMote:SOURce:CURRent <?|LOCal
|AVOLtage|IAVoltage|ARESistive|IAResistive>
[:]SYSTem[<channel>]:COMMunicate:APRogram:LEVel:CURRent
[:ISOlated]
[:]MEASure[<channel>][:SCALar]:APRogram:CURRent[:DC]?
Summary of Contents for XTR 850 Watt
Page 2: ......
Page 3: ...XTR 850 Watt Series Programmable DC Power Supply firmware v1 09 and below Operating Manual...
Page 6: ......
Page 18: ...xvi...
Page 30: ...1 8...
Page 52: ...Local Operation 3 8 M370046 01 Figure 3 3 Front Panel Menu System...
Page 126: ...Analog Programming APG and Isolated Analog Programming ISOL 4 34 M370046 01...
Page 198: ...5 72...
Page 234: ...A 36...
Page 284: ...IX 4...
Page 285: ......