Calibration
Entering Calibration Mode
Release 3.0
181
SCPI
To access calibration mode via remote interface, use the command:
CAL:STAT ON, "0000"
1
To check if the power supply is in calibration mode, use the command:
CAL:STAT ?
Security code
To protect calibration data, a security code is required to enter calibration mode. The
security code set at the factory to "
0000
."
The password can be changed from the remote interface or the front panel.
Calibration state must be ON to change the password. From the front panel, select
"
Change Code
" from the calibration menu and enter the new code.
To change the Calibration security code:
If you have entered a valid code, the prompt displays
Output V Cal
. You can
either continue working with the calibration options or press
EXIT
to leave
calibration mode.
The SCPI command to change the security code is:
CAL:CODE <security_code>
The security code is any 4-digit number enclosed by quotation marks. Trying to
change the password to an invalid one causes an error.
1. where the parameter "0000" may be replaced with your own 4-digit security code.
Step # Do This
You Will See
1
Output V Cal
2
or
Scroll to select
Change Code
.
Change Code
3
Enter a new 4-digit code.
Code ####
4
You are back in the Calibration menu.
Output V Cal
ENTER