Operation
Power Supply Operation
Release 3.0
97
Using Auto
Sequencing
Auto Sequence programs can be set to run as a Power ON default or recalled from
memory by pressing the
RECALL
key. In Auto Sequence mode, 3 of the function
keys operate as alternates:
•
VOLTAGE
operates as
RUN/PAUSE
.
•
CURRENT
operates as
TRIGGER
.
•
STORE
operates as
END
.
In the following discussion about running programs in Auto Sequence mode, each
of the keys mentioned above is referred to as their alternate function.
When an Auto Sequence program is launched from Recall, the AUTO SEQ
annunciator illuminates on the front panel. Press
EXIT
to return the unit to normal
operating mode.
Auto sequence programs can operate 3 different ways:
•
They can run automatically through a series of steps, repeating those steps a
pre-set number of times if necessary, and complete their operation without
intervention from an operator.
•
They can run automatically and be paused. If you need to stop the program
temporarily, you can press
PAUSE
to stop it, and then press
RUN
to resume the
program’s operation when ready. When a sequence is manually paused, the
Pause annunciator is illuminated. (The output remains on and voltage may be
present at the output.)
•
They can run automatically programmed to wait for a trigger at certain points in
the sequence. If you want the program to resume, you can press
TRIGGER
or
supply a trigger signal to the rear panel Trigger input. When a sequence is paused
by a trigger, the Trigger? annunciator is illuminated.
During operation, press the Up key to display information on the currently running
sequence including sequence number, step number, step parameters (settings), the
trigger source, as well as the loop count.
To run an auto sequence program:
Step # Do This
You Will See
1
4 times
Auto Sequence
2
Auto Seq 1
3
Seq 1 ready
4
#####V #####A
RECALL
ENTER
ENTER
RUN