Remote Operation
SCPI Commands for Digital Interfaces
126
Operating Manual for XDC Series Power Supply
Table 4.8
Features Affected by Reset (*RST) Command
Feature
Reset State (*RST)
Voltage setpoint
0.0V
Current setpnoint
0.0A
Power setpoint
103% of power rating
Low voltage setpoint limit 0.0V
High voltage setpoint limit 103% of voltage rating
Low current setpoint limit 0.0A
High current setpoint limit 103% of current rating
Low power setpoint
3% of power rating
High power setpoint
103% of power rating
Over voltage protection
Disabled (0.0V)
Under voltage protection
Disabled (0.0V) and not shutdown when tripped
Over current protection
Disabled (0.0A) and not shutdown when tripped
Under current protection
Disabled (0.0A) and not shutdown when tripped
Over power protection
Disabled (0.0W) and not shutdown when tripped
Under power protection
Disabled (0.0W) and not shutdown when tripped
Fold shutdown protection None and delay 0.5s
AC off shutdown recovery Autorecover
Over temperature
shutdown recovery
Latched
Triggered voltage
setpoint
Disabled (Default)
Triggered current setpoint Disabled (Default)
Triggered power setpoint Disabled (Default)
Autosequence mode
Exit
Front panel display config Show V & I
Knob lockout
None
Calibration mode
False
Output
Disabled