Remote Operation
GPIB Operation
Release 3.0
121
Configuration
Configure the power supply’s GPIB address and power-on service request setting.
The defaults are GPIB address 2 and power-on service request off.
Front panel
First set the GPIB parameters:
Next, select GPIB as the remote control interface.
Step # Do This
You Will See
1
ERROR MSGS
2
5 times
REMOTE CONFIG
3
RS-232 Cfg
2 times
GPIB Cfg
4
GPIB Addr ##
5
Select an address from 1-30
GPIB Addr 2
6
PON SRQ? Y
7
or
Select whether power-on service
request is to be sent.
Options are
Y (Yes) or
N (No)
8
Settings are saved.
Step # Do This
You Will See
1
ERROR MSGS
2
4 times
REMOTE SELECT
3
RS-232
4
or
Select GPIB.
GPIB
5
GPIB setting is saved.
MENU
ENTER
ENTER
ENTER
ENTER
MENU
ENTER
ENTER