U
SER
M
ANUAL
C
OMMISSIONING
31
4.7
Configuring switches
Configure the built-in terminators via the rotary switches on the back
of the ML-N4000.
For the LIN interface, select between internal or external power supply
via the toggle switch.
4.7.1
4x CAN
Position
Resistance
0
Open (∞)
1
2.6 kΩ
2
120
Ω
The channels 3 and 4 have the value
Open
and, therefore, are not
terminated.
4.7.2
2x FlexRay
Position
Resistance
0
Open
(∞)
1
2,6 kΩ
2
90 Ω
4.7.3
7x LIN
Position
Power supply
Vi
Internal
Ve
External
Use the internal power supply of the LIN measurement card only if the
connected LIN device uses the ML-N4000 voltage as a reference voltage.
Bear in mind that this is not possible when using the ML-N4000 in a 24-V
electrical system, for example. In this case, connect the external power
supply port to the corresponding reference voltage.