What to do if...
67
ba75958d05
10/2017
6
What to do if...
6.1
General errors
Display,
LoBat
Instrument does not
react
to keystroke
RS232 interface
does not react
Error message,
Error
0, 8, 16, 16384
Cause
Remedy
– The batteries or rechargeable
battery are largely depleted
– Insert new batteries
– Charge the rechargeable
battery (see section 3.2)
Cause
Remedy
– Software error
– Operating condition undefined
or EMC load unallowed
– Processor reset:
Press the
<START/ENTER>
and
<PRT>
keys
simultaneously.
Cause
Remedy
– Software error
– Operating condition undefined
or EMC load unallowed
– Processor reset:
Press the
<START/ENTER>
and
<PRT>
keys
simultaneously.
Cause
Remedy
– Instrument error
– Repeat measurement
– Meter defective,
send meter to WTW for repair
and quote the error number